Messages and Special Case: Gage R&R

Reading time
Message or situationCause and Solution
«Please fill in all the values»The entry table contains empty cells. The R&R requires a complete plan.
«This type of ANOVA cannot be performed with 1 repetition.»Increase to at least 2 repetitions, or choose the "averages and ranges" method.
Insufficient resolutionThe instrument does not have enough significant digits for the tolerance. No study will be able to make up for this.
Low NDC but good CPCA method suitable for compliance screening but unsuitable for detailed analysis. Acceptable depending on the application.
GRR %: excellent, very wide toleranceVerify that the tolerance is indeed the one specified in the drawing, and not a wider value.
Pieces That Are Too SimilarThe «parts» share is artificially low, and the GRR % appears poor. Include parts that cover the entire production range.